BendTec

Flexural rigidity tester

You would like to have more information?
We are at your service!

You can reach us Monday to Friday between

8:15 a.m. and 5:00 p.m.
Universal Time Coordinated (UTC) +1

by phone

+49-2166-1898-500

You are welcome to send us an email though.

For ease of use, this simple to use tester is equipped with a touch screen. The progress of the test is displayed on the screen. Charts, single and statistical values are displayed immediately at the end of the test. The robust housing is set up on specially developed vibration dampers. The acryl cover keeps air blasts and other external impacts away that might lead to incorrect results particularly when low forces are applied. Bending angle, angular velocity, dwell time and other parameters can be preselected. Norm-based testing methods are predefined. User-defined test methods can be stored.

Test description:

A specimen is set up and, in the manual version of the tester, clamped by means of a torque screwdriver. With the pneumatic version of the tester, the specimen is clamped automatically when the start button is pressed. The specimen loader releases the specimen for the measurement and the specimen is pressed into contact with the bending cutters of the force transducer. The curve is already displayed while the measurement is still in progress. When the present bending angle is obtained, the specimen stops and is held for the preset dwell time, after which it returns to the starting position.

The automatic version releases the specimen automatically. The test results and the test curve are displayed. When longitudinal and transversal tests are carried out, a ratio of both is also displayed.

Applicable norms:

  • ISO 2493
  • VDA checksheet 230-209 for leather
  • SCAN P29
  • DIN 53121
  • BS3748
  • TAPPI T556
  • NFQ03-048

Specification:

  • Accuracy +/- 1% of the scale in the range from 2 per thousand to the nominal load of the force sensor
  • Force sensor 2N, 5N, or 10N
  • Bending angle 0-92°
  • Angular velocity setting from 0.,-50°/s
  • Specimen width 38mm (paper), 30mm (leather and plastics)
  • Pneumatic clamping
  • Automatic carriage shifting available
  • 3 measuring points (angles) and a dwell time can be set individually
  • Display of individual values, diagrams, statistical functions
  • Control and measuring value processing system with Windows like touch screen operation
  • RS 232 interface for result data transfer and parallel printer interface for an HP Laserjet compatible printer
  • Compatible with MiniStat and PaperLIMS

Features:

  • Zero backlash gear
  • According to ISO and TAPPI suitable for applications with paper or according to VDA for applications in the automotive area
  • Touch screen operation
  • Display of individual values, diagrams, statistical values and ratio

Technische Daten:

Power supply 110-230V, 50-60Hz
Water nein 
Compressed air 400-600 kPa
Dimensions in cm (W)38 x (D)45 x (H)32 
Weight net 27 Kg 
Weight gross  37 kg 
File Description Date added File size
pdfBendTec Test stand bending stiffness
2.12.2015 17:15 743 KB